Product Specifications
|
Parameter |
Specification |
|
Model |
KL-8071A |
|
Pumping Mechanism |
Curvilinear peristaltic |
|
IV Set |
Compatible with IV sets of any standard |
|
Flow Rate |
0.1-1200 ml/h (in 0.1 ml/h increments) |
|
Purge, Bolus |
600-1200 ml/h (in 1 ml/h increments); Purge when pump stops, bolus when pump starts |
|
Accuracy |
+3% |
|
VTBI |
1-20000 ml |
|
Infusion Mode |
ml/h, drop/min, time-based |
|
KVO Rate |
1-5 ml/h |
|
Alarms |
Occlusion, air-in-line, door open, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, standby |
|
Additional Features |
Real-time infused volume, automatic power switching, mute key, purge, bolus, system memory, key locker |
|
Occlusion Sensitivity |
High, medium, low |
|
History Log |
30 days |
|
Air-in-line Detection |
Ultrasonic detector |
|
Wireless Management |
Optional |
|
Vehicle Power (Ambulance) |
12 V |
|
Power Supply, AC |
100-230 V, 50-60 Hz, ≤30 VA |
|
Battery |
12 V, rechargeable |
|
Battery Life |
8 hours at 25 ml/h |
|
Working Temperature |
10-30°C |
|
Relative Humidity |
30-75% |
|
Atmospheric Pressure |
860-1060 hpa |
|
Size |
15012560 mm |
|
Weight |
1.7 kg |
|
Safety Classification |
Class II, type CF |